PosiTector 6000
INSTRUCTION MANUAL v. 6.2
for Basic (1) and Standard (2) models
Simple. Durable. Accurate.
Coating Thickness Gages
Separate
probe
Built-in
probe
Przeglądanie stron 0
1 2 3 4 5 6 ... 15 16

Podsumowanie treści

Strona 1 - PosiTector 6000

PosiTector 6000INSTRUCTION MANUAL v. 6.2for Basic (1) and Standard (2) modelsSimple. Durable. Accurate.Coating Thickness GagesSeparateprobeBuilt-inpro

Strona 2 - Power-up / Power-down

SSttatistical Functionsatistical Functions-when Statistics is selected, a icon andstatistical summary will appear on the display.Remove the last

Strona 3

11-zeros all on-screen Statistics and HiLo tabulations.is above the Hi limit. The LED will blink red ifreadings are outside limits. Press (+) to clear

Strona 4

When Hi Res is selected, the displayed gage resolu-tion becomes as follows:ResolutionRange0.01 mil 0.00 - 99.00 mils0.1 mil 100.0 - 999.9 mils0.1 um 0

Strona 5

13SepSeparate Probesarate ProbesSeparate Probe Gages consist ofa gage body and a probe. A wideselection of interchangeableprobes are available. Each r

Strona 6

14automatically attempts a measurement using theeddy current principle. If the coating is non-conductive over metal, a reading is displayed withthe le

Strona 7

15TTroubleshootingroubleshootingSome common reports received by our ServiceDepartment along with possible causes are locatedon our website. Most condi

Strona 8

16www.defelsko.com© DeFelsko Corporation USA 2009All Rights ReservedThis manual is copyrighted with all rights reserved and may not be reproduced ortr

Strona 9 - Memory Mode

2IntroductionIntroductionPower-up / Power-downPower-up / Power-downThe PosiTector 6000 powers-up when any button ispressed. To preserve battery life,

Strona 10

3Quick SQuick Sttartart1.Separate probe models - remove red protectiverubber cap if supplied. Built-in probe models -remove Gage from protective rubbe

Strona 11 - Setup Menu

4Menu OperationMenu OperationCalibration,Calibration,VVerificationerificationand and AdjustmentAdjustmentThree steps ensure best accuracy…1.Calibratio

Strona 12

5VerificationVerification is an accuracy check performed by theuser using known reference standards. A successfulverification requires the Gage to re

Strona 13 - Not this!

6With “FN” Gages, calibration adjustments aremade only to the “F” or “N” mode (stored independentlyunder a particular Cal), whichever was measured las

Strona 14

7(3) Simple Adjustment to a Known ThicknessIt is sometimes desirable to adjust the Gage to aknown thickness, such as a shim, rather thanadjusting it t

Strona 15

8thicker value. These values should be on either sideof the thickness range to be measured.1.Select 2 Pt Adjust from the Cal Settings menu.2.Press (+

Strona 16

9Memory ModeMemory ModeStandard (2) Gages onlyThe PosiTector 6000 can record 250measurements for on-screen statistical purposes, forprinting to an op